CARE AND USE OF THE MICROSCOPE
After completing this investigation, you
should be able to:
1.
Identify the various parts and associated functions of your microscope.
2.
Know how to properly adjust the light sources to achieve optimum illumination.
3.
Take advantage of the parfocal nature of the objectives.
4.
Understand the relationship between magnification and size of the field you are
viewing.
5.
Demonstrate your ability to find and focus on an image at all powers.
Introduction
The bright-field compound
microscope is probably the most important instrument you will use in any
biology course. Objects such as bacteria, mold spores, plant cells, or animal
tissues, too small to be seen with the unaided eye will become highly
magnified, allowing you to view minute details.
Logic would suggest
that the more we magnify the image of a cell, the more detail we can see.
Unfortunately, this is not true. Light rays passing through curved surfaces are
bent. This bending of light produces an enlarged image. But the best research
microscopes can currently magnify only 1500 to 2500 diameters. The limiting
factor is the nature of light. Of equal importance with magnification is resolving power. Resolving power or
resolution is the ability to distinguish separate images of objects that fall
very close together on the retina of the eye. Since the human eye cannot
ordinarily resolve objects separated by less than 0.1 mm, a microscope can
extend the resolving power of the eye about 400 times. Thus, a microscope
provides magnification and permits
us to see objects as separate images that are close together and would appear
as a single object to our eyes.
Parts of the Compound Microscope (relate the information to
Figure x-1)
1. The microscope
rests upon the table on its base or foot.
2. Attached to the foot is the upright part
known as the arm or pillar of the microscope.
When carrying the microscope grasp it
firmly by the arm. DO NOT TIP IT. If
it is too
heavy, support the microscope by holding
one hand under the base.
3. The stage
is the platform upon which the slide to be studied rests. Our microscopes have
mechanical stages which enable you to
easily control the movement of the slide you are
observing. Below or next to the stage,
usually on the right hand side, are two milled knobs.
They control the movement of the mechanical stage. The inner knob will
move the
stage toward or away from you. The outer
knob gives a right-left motion to the slide holder.
To place a slide on the stage, pull
back the lever arm that protrudes from
the left side of the
mechanical stage. This opens the slide
holder and allows you to insert the slide into the notched
area. Release the lever carefully to
avoid damage to the slide.
4. The coarse
adjustment is the larger milled wheel located near the base of the
microscope, on each side. This moves the
stage up and down and enables you to get an
approximate focus on the study object.
About 90 –95% of your focusing is done with the coarse
adjustment.
5. The fine
adjustment is the smaller milled wheel at each side of the base of the
microscope. This adjustment moves the
stage up and down small distances to bring objects into
sharp focus. You should be able to use
this adjustment to gain knowledge about the relative
position of the parts of the object you
are observing. Focusing up and down will bring new
parts into clear focus, and tell you
that these parts are above or below the parts on which you
had previously focused. Remember that
even though small, objects you are observing are three-
dimensional. Thus, they require frequent
readjustments of focus. This will aid in development
of a clearer concept of the structural
relationships between the parts of the object being viewed. If,
when attempting to use the fine
adjustment, you find that the wheel has been turned as far as it will
go in one direction, turn the coarse
adjustment so that the lenses are farther from the object
being viewed. Focus again with the
coarse adjustment and then turn the fine adjustment. If you
have correctly focused with the coarse adjustment, you will
not need to turn the fine
adjustment more than one half of a turn.
6. The revolving
nosepiece is at the lower end of the body tube. It normally holds two to
four objectives. Turn the coarse adjustment so that there is room for
the longest adjustment
without it touching the stage. Now turn
the nosepiece and feel the "click" as each objective
goes into place.
Some microscopes
have a very short objective which magnifies about 4 times. This is sometimes
called the scanning objective and
can be used to get a wide view of the field of study. The next shortest
objective has a large opening in it and a large lens. This objective is called
the low power objective and magnifies about 10 times. In some models of
microscopes the number 16 is inscribed in this objective. This indicates that
the objective is 16 mm above the object when the object is in focus. The high-dry objective is longer than the
10 power objective and is identified by the number 40 or 45 inscribed on it. It
should magnify about 40-45 times.
The microscopes in
many laboratories, ours included, have an oil
immersion objective. Note the size of the aperture (opening) in this lens.
To use this lens place a drop of immersion oil on the slide or cover slip.
Place the oil exactly where the light is coming through the opening in the
stage and onto the slide. The lens is then carefully moved into the oil forming
a continuous column from the slide to the objective lens. Light is not
refracted upon passing between the oil and the glass slide. The oil immersion
objective magnifies about 100 times.
Immersion oil
contributes to finer resolution and brightness. These characteristics are most
critical under higher magnification. Thus, only the higher power, shorter focus
objectives are designed for oil immersion.
For any given lens
there is a fixed focal length. With
the objective in focus there is a cone of light extending from a point on the
specimen to the full diameter of the objective lens. Below the specimen is a
second, matching cone of light. The base of the cone is the top surface of the condenser and the apex at a point on
the specimen. Theoretically, illumination provides a straight line path for
each ray from condenser to objective lens. Immersion oil contributes to the
homogeneous path of light between the objective and the condenser by having the
same index of refraction as the glasses in the system.
These objective
lens are parfocal. This means that
they are designed so that once you have brought an image into focus under low
power, you may switch safely to one of the higher powers without hitting the
slide or coverslip or making major readjustments with the coarse adjustment.
You will find that the object you are viewing will be in approximate focus. If,
when switching to a higher power, you meet resistance, check the slide. You may
have an unusually thick coverslip or you may have your slide on the stage
upside down. Before you move to a higher power, be sure to place the object to
be viewed in the center of the present field of view. Remember that as you increase magnification you decrease proportionately the diameter
of your field of view.
7. At the upper end of the tube are two oculars or eyepieces. Each contains two
lenses. Note
that they do not fit tightly into the body
tube. This is one of the reasons you are asked to carry
the microscope in an upright position.
Note that each ocular bears the marking 10X or 10 on
the side. This means that it magnifies
10 times. This ocular number, multiplied by the number
on the objective, gives the approximate
magnification obtainable with that objective.
It is very
important to locate dirt particles on the lens system and to remove them with
lens paper or lens paper moistened with alcohol or glass cleaner BEFORE beginning the laboratory work.
If a blur appears, a sharp image of the object cannot be obtained until the
cause is corrected. As you look through the oculars, turn them. If the smear
turns with the ocular, then the ocular is dirty. If it does not turn, try
moving the slide. If the dirt is still present, switch the objectives and see
if the dirt is on the objective. If not located at any of these points, clean
the condenser and/or the built-in light source. In general, clean all surfaces
through which light passes.
8. Beneath the stage are a series of lenses
which serve to bring the light rays into focus. Known as
the condenser,
it may be raised or lowered by turning the milled knob under the left side of
the stage (models will vary). To focus
the condenser, turn the knob of the light source to a low
setting. Note that there is only a small
spot of light coming from the light source. Put the
microscope on low power (10X objective)
and focus the objective lens on the object on the
slide. Raise or lower the condenser
until a sharp hexagonal shape can be seen through the
ocular. The condenser is now in focus
for this particular slide. In general, moving the condenser
lens away from the stage causes light to
be diffused or spread out. Moving the condenser toward
the stage brings the light to a point on
the slide. As you increase magnification you will normally
have to move the condenser up towards
the stage. Remember, as you increase magnification
you
also must increase illumination.
9. Beneath the stage of the microscope,
below the condenser, and sometimes attached to the
condenser is a series of thin metal
leaves which can be opened or closed to regulate the amount
of light passing through the lens system
of the microscope. This is the iris
diaphragm.
The iris diaphragm is opened and closed
by moving the lever found under the front-right part of
the stage. You should adjust this every
time that you look at a new object. This will ensure that
you are achieving maximum contrast.
10. The light source is built into the base
of the microscope and controlled by a rheostat knob or
lever.
Proper Use of the Compound Microscope
1.
Clean all lens surfaces with lens paper and place the object to be viewed on a
slide on the stage
of the microscope. Be sure that the 10X
objective is in position.
2. Adjust the light coming through the
microscope so that the object is evenly illuminated and the
brightness allows for comfort without
squinting.
3. Without looking through the microscope,
bring the stage up towards the objective until it stops.
The microscope should have a built-in
stop to prevent damage to the lens system at this
magnification. Move the slide so that
the object you are viewing is centered in the circle of
light coming through the opening in the
center of the stage. If no light is present, check to be
sure the light source is on, the plug is
in the electrical outlet, and the diaphragm is open
4. Look through the oculars as you turn the
coarse adjustment and lower the stage until the object
is in focus.
5. Adjust the iris diaphragm and condenser
for good lighting.
6. Remember that the objectives on these
microscopes are parfocal. Before moving the high
power lens into position, move the
object to view into the center of the field.
Precautions to be Observed when using
the Compound Microscope
1. Never touch any of the lens with anything
other than lens paper and approved lens cleaner.
2. Always rack down to bring the object into
focus.
3. Keep both eyes open at all times.
4. All initial
focusing is done with the coarse
adjustment. Use the fine adjustment to sharpen
your final focus and to note the
three-dimensional nature of the object in view.
5. Always clean the lenses before and after
using the microscope. Be sure to clean all oil from the
lens after final use.
6. Before returning the microscopes to the
cabinet, place the lowest power objective in place. Be
sure the mechanical stage is not
sticking out and be sure that the power cord is not severely
bent at the point where it leaves the
base. Recover the microscopes before replacing them in the
cabinet.
Add
picture of binocular microscope Figure 1. Compound
Microscope G045