OPERATOR SELF-TEST OF 5004A
The 5004A Signature Analyzer has a SELF-TEST mode which can be used to check the condition of the unit thoroughly. Use the following procedure to SELF-TEST a 5004A.
Before applying power to the 5004A check that the line (mains) voltage available matches the label on the 5004A rear panel.
Remove the grabber connectors from the pod test leads, and connect the pod (START, STOP, and CLOCK) leads to the matching START, STOP, and CLOCK receptacles on the 5004 front panel.
Connect the 5004A data probe to the PROBE TEST receptacle until the point is held securely.
Connect the 5004A power cable to the correct power source and set the 5004A front panel as follows:
Switch Settings |
Displays |
|||||
START |
STOP | CLOCK | Four Seven-Segment | GATE | UNSTABLE SIGNATURE | PROBE TIP LIGHT |
in | in | in or out | UP73 then ACA2 | flickers | Flickers except when good signature is on | Flickers when "ACA2" is on |
out | out | in or out | 3951 then 2P61 | flickers | Flickers except when good signature is on | Flickers when "2P61" is on |