Valve A open, B closed -- outside water flowing through livewell
Valve A closed, B open -- recirculating livewell water only
Valve A closed, B closed -- strainer D can be cleaned
Valve A open, B open -- livewell will drain
Check valve C prevents back flow
|
KEEPALIVE INSTRUCTION AND INSTALLATION MANUAL <>< <>< <>< PLEASE READ |
||||||||||
|
INDEX 1 |
||||||||||
|
POWER CONSUMPTION Battery...........13.0 Volts
Depth Air Valve Open
Air Valve Closed _______________________________________ COMMON INSTALLATION PROBLEMS 1. Pump operates but no bubbles.
2. Suction cups do not stick to bottom of tank.
3. Does not appear to operate in fresh water.
4. Pump running does not pump water.
2 |
||||||||||
|
REMEMBER THIS IMPORTANT TIP Control the water current in your container with the AIR CONTROL VALVE. The more air added with your valve, the slower your water current will be. Adjust water flow with the AIR CONTROL VALVE, until the fish are schooling and in a stationary position in the tank. Adjusting the water current for the species of your fish will keep them in prime condition. If water flow is too fast, fish will tire quickly. REMEMBER..... 4 |
||||||||||
|
OPERATION OF THE KEEPALIVE AIR CONTROL CENTER VISUAL DIFFERENCES BETWEEN SALTWATER AND FRESH WATER: In saltwater, bubbles are readily seen due to the illumination of the salt. In freshwater, almost nothing can be seen. The dust sized bubbles are extremely hard to see. Put your hand in the water and watch the millions of micro-fine bubbles gathering on your skin. Foam on the surface is fish waste, and can be skimmed off. Foam does not interfere with the KeepAlive aeration process. Control your water flow with the Air Control Valve.. Adjust the speed of water with the Air Control, so fish school in a stationary position. Opening the valve produces larger bubbles as well as the micro-bubbles, and slows the water flow. Closing the valve will increase water flow and the larger bubbles begin to disappear. Each fish species will have their own swimming speed, so water flow must be adjusted for each species. For a better understanding, please read our LESSON IN AIR BUBBLES 5 |
||||||||||
|
IMPORTANT INSTRUCTIONS FOR FRESH WATER PLEASE READ! This product works equally well in both fresh and salt water. You may have seen a picture or display of a unit in salt water producing large numbers of visible bubbles. KeepAlive, operating in salt water, provides a much more dramatic VISUAL display than fresh water. To ensure that your Infusor is operating properly, LISTEN for the sound of air at the Air Control Valve. IN FRESH WATER, the Infusor produces many sizes of bubbles. The larger bubbles, although highly visible, do very little in oxygenating the water. They rise too quickly. The Infusor produces millions of dust size bubbles that remains suspended in the water longer and are virtually invisible to the naked eye, unless you place a strong light on the water and observe from the side, as in an aquarium. 6 |
||||||||||
![]() A SIMPLE PROCEDURE TO ENSURE THAT YOUR UNIT IS OPERATING PROPERLY, IS TO PLACE YOUR HAND IN THE WATER AND WATCH THE BUBBLES COLLECT. REMEMBER THAT THE BROWN LEAD GOES TO THE POSITIVE (+) SIDE OF THE BATTERY. 7 |
||||||||||
|
INSTRUCTIONS FOR ALL KEEPALIVE PUMPS AND ACCESSORIES Operation of the KeepAlive Air Inlet Valve: When your unit is provided with a 1/8" air check valve, install it in line next to the Air Control Center with a small portion of the air hose provided. It is recommended that all threaded fittings must have no less than pipe thread tape sealant. ALWAYS DOUBLE CHECK FOR LEAKS! 8 |
||||||||||
|
THRU-HULL MODELS To install the thru-hull models we recommend you follow the livewell diagram and instructions on pages 11 and 13 as a general guide. There are too many different hull and livewell designs for only one type of installation. KeepAlive components were designed for the angler to create a custom livewell. Tips
9 |
||||||||||
|
No-LokTM INSTALLATION No-Lok Models with pump must be tilted at a 45° angle to allow air to escape by using KA105 elbows with pump output and No-Lok noozzles facin upward for the No-Lok to work properly (See Figure 1). Next, install the No-Lok relief valve valve. Frill a 1 1/16" hole above the water line in the livewell as shown in Figure 1. Slide the thru-hull in the hole and thenthread nut on ¾ of the way. Hold the No-Lok relief valve until threads tighten into the relief valve. Tighten the thru-hull nut and connect ¾ I.D. relief hose barb, then to No-LOk barb on processor. Use staniless steel hose clamps. The No-Lok is now fully operational. 10 |
||||||||||
![]() Figure 1 Valve "A" opened, "B" Closed - Outside water flowing through well. Valve "A" closed, "B" opened - Recirculating water only. Valve "A" opened, "B" opened - Livewell will drain. Valve "A" closed, "B" closed - Strainer "D" cam be cleaned. Check valve "C" prevents back flow. 11 |
||||||||||
|
CONVERSION KIT INSTALLATION
1. Remove livewell
pump. 13 |
||||||||||
RECIRCULATING KIT INSTALLATION![]()
1. Pump must be tilted at a 45°
angle on completed installation so all air can escape upwards. 14 |
||||||||||
TANK INSTALLATION![]()
1. Adjust the lid twist locks to
the desired tension. BROWN LEAD IS POSITIVE! 15 |
||||||||||
|
OXYGEN REGULATOR
The American Healthcare FISH-FLO2 Oxygen Regulator has been designed to provide a constant flow of oxygen from the cylinder, directly to the aerator or oxygen stone. To increase the oxygen flow, turn the selector knob from "0" to a setting of "1" through "11". Each oxygen regulator is equipped with an internal pressure relief valve that will prevent over-pressurization of the regulator. The American Healthcare FISH-FLO2 Oxygen Regulator is specifically designed for oxygen service only. Use of this regulator for gases other than compressed oxygen up to 2500 PSIG, is prohibited by American Healthcare and any user must assume all liabilities and consequences.
MAINTENANCE AND CARE Wipe down with a clean, soft cloth. If contaminated, the regulator should be taken out of service immediately. If your FISH-FLO2 Oxygen Regulator is accidentally immersed in salt water, remove the red dust cover from the safety relief valve, flush well with fresh water, and let dry before using again. CONNECTION GUIDELINES Never stand in front of a cylinder outlet or regulator when operating the cylinder valve. Remove the two protective covers from the 540 Valve Nipple and the Hose Barb before attempting to connect to a cylinder. DO NOT REMOVE THE 5/8" DIAMETER RED DUST COVER FROM THE SIDE OF THE REGULATOR NEXT TO THE HOSE BARB. IT IS THERE TO HELP PREVENT CONTAMINATION. REFER TO THE SAFETY TIPS AND WARNINGS BEFORE PROCEEDING FURTHER. Be sure to properly inspect both the cylinder and regulator for any signs of grease and oil. If grease or oil is suspected or found, do not use this device.
|
||||||||||
|
Do not use organic-based thread sealant on any part of this regulator. This oxygen regulator is equipped with an internal relief valve designed to relieve excess pressure should the regulator malfunction. This relief valve is designed to operate only as a means of a warning signal. Should a malfunction occur, turn off the cylinder valve immediately and disconnect the regulator. Do not attempt to operate the oxygen regulator, unless you are properly trained. CAUTION: Oxygen is a non-flammable gas, however, the presence of oxygen will drastically increase the rate of heat and combustion. Inspect for traces of oil, grease or any other type of contamination, or connection damage before operating. Failure to comply with these safety precautions can result in personal injury or property damage. Do not proceed to use until all equipment and components are correct for safe oxygen use. CONNECTING A CGA 540 NUT AND NIPPLE INLET 1. Check to be sure that
the valve on the cylinder is off. OPENING/CLOSING THE CYLINDER VALVE 1. Before turning on the
oxygen cylinder valve, check to be sure that the regulator is in the "0"
position. LIMITED-LIFETIME REPLACEMENT WARRANTY Because or the harsh unforeseen conditions that the FISH-FLO2 Oxygen Regulator may be subjected to, the regulator is covered by a limited-lifetime warranty. If a problem or defect is found at any time and caused by any reason including, but not limited to, accidental damage and contamination, return the rhe regulator to American Healthcare and receive a new replacement FISH-FLO2 Oxygen Regulator for $75.00 plus shipping. To qualify, all regulator parts must be included with the regulator and the the serial number must be legible. You must first call American Healthcare at 1-888-567-7733 and give the customer service representative a valid credit card number and obtain a Replacement Regulator Authorization Number (RRA). This RRA number must appear on the outside of the box containing the returned regulator. Any shipments received without an RRA number will be refused. A new replacement FISH-FLO2 Oxygen Regulator will be shipped out the next business day. UNDER NO CIRCUMSTANCES SHOULD ANY REPAIRS, OR REPLACEMENT OF ANY PARTS BE PERFORMED BY ANYONE EXCEPT American Healthcare. 16 |
||||||||||
|
KeepAlive
Aerators and Livewells |
| HOME | PORTABLE UNITS | RE-CIRCULATING KITS | THRU-HULL UNITS | CONVERSION KITS | |
| TANKS 10 20 & 30 GALLON | ACCESSORIES | GUIDE TO AERATION | INSTALLATION | MOVIE DEMOS | |
SEND EMAIL to FDNYFIRE@aol.com